HART Level Sensor

HART Level Sensor

▲ Measuring range:0~2.67kPa...2500kPa.
▲ Intrinsic safety and flame proof option.
▲ For level, pressure, density measurement of high temperature, macro viscosity, easy crystallized, easy precipitated or strong corrosive media.
▲ Adjustable range: 6:1 for analog type, 15:1 for intelligent type.
▲ High accuracy, good reliability and stability.
Chat Now

Product Details

HART Level Sensor


Description of HART level sensor

ULB3351 flange type HART level sensor is one kind of scene transmitter for mounting in pipelines or vessels. Because its isolating diaphragm contacts liquid medium directly, not need pressure pipe to draw out the pressure from positive pressure side (H side). ULB-3351 is suitable for level, pressure, density measurement of high temperature, macro viscosity, easy crystallized, easy precipitated media.

ULB3351 flange type HART level sensor includes plane type and insert type. The mounting flange has 3” and 4” according to ANSI standard. 

Features of HART level sensor

▲ Measuring range:0~2.67kPa...2500kPa.

▲ Intrinsic safety and flame proof option.

▲ For level, pressure, density measurement of high temperature, macro viscosity, easy crystallized, easy precipitated or strong corrosive media.

▲ Adjustable range: 6:1 for analog type, 15:1 for intelligent type.

▲ High accuracy, good reliability and stability.

Specifications of HART level sensor

pressure medium liquid compatible with stainless steel
measuring ranges0~2.67kPa...2500kPa
output signal4~20mA (HART optional)
accuracy0.1%FS,0.2%FS(standard), 0.5%FS
load resistance  RL=(U-12V)/0.02A (4~20mA current output) U—loop voltage(V)
supply voltage12~45VDC
operating temperature range-40~+85℃
measured media temperature range-40~+200℃
temperature coefficient of zero0.1%FS/10℃
temperature coefficient of span0.05%FS/10℃
process connection3" mounting flange or others
protectionIP67
explosive-proofExia II CT6 or Exd II CT6
material of pressure membrane316L stainless steel
oil-filledsilicone oil or fluorocarbon oil
Application

The ULB3351 flange type HART level sensor is especially suitable for level, pressure and density measurement of special media (high temperature, macro viscosity, easy crystallized, easy precipitated, strong corrosion) in open or sealed containers.

Inquiry